Supply |
Conditions |
Min. |
Typ. |
Max. |
Units |
|---|---|---|---|---|---|
|
Supply power |
|||||
| 12 | - | 28 | Vdc | ||
|
Power consumption |
|||||
| - | 0.9 | 1.3 | W | ||
| Power on Current | During power on | - | 130 | - | mA |
|
Fuse slow |
auto reset |
- |
0.2 |
- |
A |
|
Input Volt |
Conditions |
Min. |
Typ. |
Max. |
Units |
|---|---|---|---|---|---|
|
No. Inputs |
Isolated to bus |
- |
3 |
- |
pcs |
|
Range |
- |
-200 |
- |
+819 |
°C |
|
Wire systems |
3-wire or 4-wire |
- |
- |
- |
- |
| Max. wire resistance | 4-wire measurement 3-wire measurement (wire resistances must be equal within 0.03 Ohms for 0.1°C accuracy) |
- | - | 100 30 |
Ohms Ohms |
|
Resolution |
lowest sample rate |
- |
0.025 |
- |
°C |
|
Accuracy |
Ambient temperature 25°C, lowest sample rate |
- |
0.05 0.1 |
0.1 |
°C |
|
Non-linearity |
- |
- |
- |
0.01 |
°C |
|
Impedance |
- |
- |
10 |
- |
M Ohm |
|
Excitation current |
- |
- |
0.2 |
- |
mA |
|
Over voltage protection |
- |
- |
60 |
- |
Vdc |
|
Line rejection |
50/60 Hz |
80 |
- |
100 |
dB |
|
Temperature drift |
- |
1 |
6 |
10 |
ppm/°C |
| Sample interval times / resolution | ||||
| Sample rate config value | 1 input | 2 inputs | 3 inputs | effective resolution |
| 0 - 2 | 10 ms | 20 ms | 30 ms | 0.4°C / 0.8°F |
| 4 - 6 | 20 ms | 40 ms | 60 ms | 0.1°C / 0.2°F |
| 8 - 10 | 33 ms | 67 ms | 100 ms | 0.025°C / 0.05°F |
| 12 - 14 | 40 ms | 80 ms | 120 ms | 0.025°C / 0.05°F |
| Moving Average filter | Full-Scale step response time | ||
| Filter setting | samples averaged | 1 input @ highest sample rate | 3 inputs @ lowest sample rate |
| 0 | 1 | 10 ms | 120 ms |
| 1 | 2 | 20 ms | 240 ms |
| 2 | 4 | 40 ms | 480 ms |
| 3 | 8 | 80 ms | 960 ms |
| 4 | 16 | 160 ms | 1920 ms |
|
Connectors |
Type |
|---|---|
|
Power supply |
2 pole screw/plug connector |
|
Input 1-3 |
4 pole screw/plug connector (Gnd, InL, InH, Vex) |
|
Cdios bus |
10 pole Flatcable female |
|
10 pole Flatcable male |
|
Signaling LEDs |
Color |
Text |
|
n.a. |
n.a. |
n.a. |
|
General |
Conditions |
Min. |
Typ. |
Max. |
Units |
|---|---|---|---|---|---|
|
I/O module |
Controlled by Cdios controller |
- |
- |
- |
- |
|
Module select |
via 4 dipswitches |
- |
- |
- |
- |
|
MPU |
Motorola 68HC908 |
- |
8 |
- |
Mc |
|
MPU watchdog |
- |
- |
52 |
- |
ms |
|
MTBF |
T amb..=40°C |
- |
87000 |
- |
h |
|
Warranty |
- |
- |
2 |
- |
year |
|
Environmental |
Conditions |
Min. |
Typ. |
Max. |
Units |
|---|---|---|---|---|---|
|
Operating temperature |
- |
-20 |
- |
70 |
°C |
|
Storage temperature |
- |
-25 |
- |
80 |
°C |
|
Relative humidity |
at 40°C non Cond. |
- |
- |
90 |
% |
|
Burn in cycle |
25°C .. -25°C .. 60°C |
- |
1 |
- |
cycle |
|
Physical |
|
|---|---|
|
Housing |
35 mm DIN rail |
|
Dimensions L * W * H |
96 * 61 * 55 mm (incl. rail) |
|
Color |
Light Grey |
|
Protection class |
IP32 |
|
Weight |
104 g |
|
Conformity |
|
|---|---|
|
EN 50081-1 |
Electromagnetic compatibility -Generic emission standard- Residential, commercial and light industry environment |
|
EN 50081-2 |
Electromagnetic compatibility -Generic emission standard- Industrial environment |
|
EN 50082-1 |
Electromagnetic compatibility -Generic immunity standard-Residential, commercial and light industry environment |
|
EN 50082-2 |
Electromagnetic compatibility -Generic immunity standard- Industrial environment |
|
EN 55022 |
Emits and methods of measurement of radio disturbance characteristics of information technology equipment |
|
EN 61010-1 |
Safety requirements for electrical equipment for measurement, control and laboratory use |
|
Last Modified: 8 February 2001 |